Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00867-1
Product Name/Description Approach Micro Wire Guide Devices

Hydro ST Microwire Wire Guide - Ref part #: HMW-14-190-ST, lot #: 15849644

CTO Microwire Wire Guide - Ref part #: CMW-14-300-18G, lot #’s: NS15811659, NS15898544, NS15898545

ARTG 224864
(William A Cook Australia Pty Ltd - Peripheral vascular guidewire)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/10/2024
Responsible Entity William A Cook Australia Pty Ltd
Reason/Issue Cook Medical has identified that devices from the affected lots have labels that state the incorrect expiration dates.

During a review of label prints, lots with labels stating incorrect expiration dates were detected due to a calculation error in Cook’s labelling software, which uses a calculator to determine the correct expiration date. The calculator assigned the maximum potential shelf life to lots in scope rather than the actual shelf life.
Recall Action Recall
Recall Action Instructions Customers should examine inventory immediately to determine if they have affected product.

If so, customers should immediately quarantine any affected products and cease all further distribution and use of the affected products.

Customers can return the affected products to Cook Medical to receive a product credit.
Contact Information 1800 777 222 - Cook Medical Customer Relations