Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00821-1
Product Name/Description Check-Flo Performer Introducer Set

Reference part number: RCF-5.0-35-45-J

Order number : G27242

Lot number: 15506769

UDI: (01)00827002272422(17) 260626(10)15506769

Expiry date: 23-Jun-2026

ARTG 137859
(William A Cook Australia Pty Ltd - Standard Introducer Sets - Catheter introducer)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/09/2024
Responsible Entity William A Cook Australia Pty Ltd
Reason/Issue Devices from lot 15506769 may have white caps that connect the shaft to the hub of the products that may disconnect. There is a possibility that this may result in an increased procedural time, harms associated with device fragmentation/separation, and/or haemorrhaging/blood loss.
Recall Action Recall
Recall Action Instructions Customers are to examine inventory immediately for affected product, quarantine and cease use. Once returned to Cook Medical, a product credit will be issued.
Contact Information 1800 777 222 - Cook Medical Customer Relations