Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00643-1
Product Name/Description BD Phoenix M50 instrument. An in vitro diagnostic medical device (IVD)

Catalog Number: 443624

All Serial Numbers

ARTG 194808
(Becton Dickinson Pty Ltd - Instrument/analyser IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 29/07/2024
Responsible Entity Becton Dickinson Pty Ltd
Reason/Issue BD identified through post-market surveillance an increase in the number of complaints for Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix M50 instrument. This problem impacts customers who are using the system for identification.
Recall Action Product Defect Correction
Recall Action Instructions Customers are to:
1. Immediately review the instrument and confirm the software version installed.
2. Complete the Acknowledgement Form in the letter and send back to anz_quality@bd.com stating the software version.
3. BD will book a service visit to provide an update to the PUD (Phoenix update data) and software to correct this problem.
4. Communicate this notification with appropriate members of their facility.
Customer can continue use of the instrument; however, confirmatory testing should be performed on any suspected E. coli misidentifications.
Contact Information 1800 656 100 - Becton Dickinson