Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00635-1
Product Name/Description HMC feeding, aspiration and feeding/aspiration tubes

All product codes beginning with: LG, NGP, NJP, RT, ST and EF2EL

ARTG 276654
(Intermed Medical Pty Ltd - Tube, feeding, nasoenteral)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 29/07/2024
Responsible Entity Intermed Medical Pty Ltd
Reason/Issue Erasure of the graduation markings on tubes.

The problem was observed to be due to the use of incorrect cleaning and disinfection processes and excessive rubbing on the outer surface of the tubes.
Recall Action Product Defect Correction
Recall Action Instructions The following indications are being added to the relevant Instructions for Use (IFU):

If necessary, clean the device with gentle movements avoiding excessive rubbing, and/or using neutral nonaggressive
detergents.

Updated IFUs will be made available with the next batches of manufactured products.
Contact Information 1300 123 633 - Intermed