Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00565-1
Product Name/Description BioPince, BioPince Ultra and Tru-Core II Automatic Biopsy Instruments

Multiple products and lot numbers

ARTG 223223
(Culpan Medical Pty Ltd - End-cut biopsy gun, single-use)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 15/07/2024
Responsible Entity Culpan Medical Pty Ltd
Reason/Issue Holes have been identified in the packaging trays which may compromise the sterile barrier of the device.

Non-sterile product potentially exposes patients to the introduction of micro-organisms into the access site, leading to an infectious process, bacteraemia, or sepsis. There is also a risk of cross contamination if the device is used with another product or on a sterile field.
Recall Action Recall
Recall Action Instructions Customers are to:
- Inspect and quarantine all impacted stock to prevent further use.
- Return affected stock for a replacement.
Contact Information 1800 060 168 - LifeHealth Care