Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00556-1
Product Name/Description Alinity c Proteins Standard. An in vitro diagnostic medical device (IVD).

List Number: 08P6701

Lot Number: 01594L823

Expiry Date: 31/05/2025

UDI: (01)00380740151324(17)250531 (10)01594L823

ARTG 205773
(Abbott Australasia Pty Ltd Diagnostic Division - Clinical chemistry-specific protein IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/07/2024
Responsible Entity Abbott Australasia Pty Ltd Diagnostic Division
Reason/Issue Some customers have experienced Alinity c Proteins control values out of range high when calibrating the Alinity c A1- Antitrypsin Reagent Kit.
This problem has not been observed when calibrating the Alinity c A-1-AGP Reagent Kit.

The problem observed may cause a delay in generating patient results if the controls report out of the acceptance range.

To date, there have been no reports of adverse events related to this problem.
Recall Action Recall
Recall Action Instructions · Customers are to stop using Alinity c Proteins Standard lot 01594L823 and discard any remaining kits of lot 01594L823.
· Arrange with Abbott to obtain a refund for affected stock.
· Replacement inventory is available to order.
Contact Information 1800 816 696 - Abbott Customer Support