Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00498-1
Product Name/Description Whole-Body MR scanners

Multiple product names, software versions and GTIN's

ARTG's: 223115 & 231238
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/06/2024
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met:

1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected,
AND
2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration.

If this occurs, it can result in overheating of an MR conditional implant.

There have been no injuries reported to GE HealthCare as a result of this problem.
Recall Action Product Defect Correction
Recall Action Instructions Users may continue to use the MR system by following the instructions outlined in the customer letter.

A GE HealthCare representative will contact all affected customers to arrange for the correction at no cost.
Contact Information 1800 659 465 - GE HealthCare Service