Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00463-1
Product Name/Description Incisive CT System

REF: 728143, 728144

Device Identifier: 00884838085015, 00884838105508

Multiple Serial Numbers

ARTG 352131
(Philips Electronics Australia Ltd - X-ray system, diagnostic, computed tomography, full-body)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/06/2024
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue A hardware problem with a metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system has been identified. This component’s structural integrity may become compromised, resulting in potential contact with other gantry components during rotation. Other components could be damaged due to the contact with the detached component.

If this problem occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs:
· Damaged component(s) come loose
· The damaged component(s) contact other components in the system and cause them to break
· A broken component makes contact with the top right gantry cover

The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled, that could cause injury, such as pain (body or neck), laceration, or eye injury. If the Incisive CT system shuts down during clinical scanning, a rescan may be needed.
Recall Action Product Defect Correction
Recall Action Instructions Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue.

Customers can continue to use the system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear.
Contact Information 1800 251 400 - Philips Service Delivery Team