Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00290-1
Product Name/Description Rad-G Device and Rad-G Device with temp

Part numbers: 27977 and 27978

Multiple serial numbers

ARTG 138873
(Masimo Australia Pty Ltd - Oximeter, pulse)
Recall Action Level Consumer
Recall Action Classification Class I
Recall Action Commencement Date 22/04/2024
Responsible Entity Masimo Australia Pty Ltd
Reason/Issue Rad-G devices are powering off and on without pressing the power button. Masimo’s investigation identified a problem that can result in an unintentional change in the power state.

Masimo has received a total of 529 complaints related to this issue. Masimo has received two reports of patient deaths while being monitored with Rad-G devices. If the device powers off unexpectedly, it could result in a loss of monitoring, which could potentially result in a delay in patient care.
Recall Action Recall
Recall Action Instructions Customers are to:
- Inspect and quarantine any affected devices to prevent further use.
- Contact Masimo Technical Services to request return of the affected product for repair or replacement (more details provided in the customer letter supplied by the Sponsor).