Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00288-1
Product Name/Description Surdial X

Serial number: 22DS10671

ARTG 211174
(Nipro Australia Pty Ltd - Haemodialysis system, institutional)
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 11/04/2024
Responsible Entity Nipro Australia Pty Ltd
Reason/Issue Nipro Australia has supplied dialysis machines Surdial X with European type C power plugs. An electrical fault on Surdial X machine was reported in which the device alarmed and the plug was blackened.

Nipro are replacing all incorrect plugs with the Australian type I plug that meets the Australian standards.

This recall action was undertaken prior to TGA notification.
Recall Action Product Defect Correction
Recall Action Instructions Nipro advise that they will replace the European type C plug with an Australian type I plug purchased in the Australia market that has met Australian standard, and electrical safety tests have also been conducted as part of the installation process.

This action has been closed out on 08/05/2024