Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00262-1
Product Name/Description Weislander Self Retaining Retractor 14cm BL

Part number: ENG-11-2033-00

Lot number: WO382835

ARTG 172560
(Device Technologies Australia Pty Ltd - Retractor)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/04/2024
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue Device Technologies Australia Pty Ltd (DTA), in conjunction with the manufacturer, Bolton Surgical Ltd, are issuing this Product Defect Correction for the product code and lot number above to the possibility for the retractor teeth to contain metal burrs.
Recall Action Product Defect Correction
Recall Action Instructions Action Required:
1. Inspect the retractor teeth of Weislander Retractors carrying the affected lot number (the lot number is contained in the barcode).

2. If no burrs are found on the retractor teeth the retractor(s) can continue to be used.

3. If any of the retractor teeth appear to contain burrs, quarantine the affected retractor(s).

4. Take a photo showing the burrs on the retractor and include this with the completed Customer Acknowledgement Form to Device Technologies.

a. Complete a form even if you have no affected devices.
b. Do not dispose of affected devices until Device Technologies confirms to do so.

This action has been finalised on 12/7/2024.
Contact Information 0490 853 258 - Jessica Hill, Product Manager