Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00176-1
Product Name/Description DX-D100 Mobile

Serial Numbers: A5411001012, A5411001212, A5411001213, A5411002010, A5411002168, A5411003095, A5411003118,
A5411003175, A5411003176, A5411003177, A5411003178, A5411003262, A5411003317

ARTG 178856
(Agfa Australia Pty Ltd - X-ray generator, diagnostic, mobile)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/03/2024
Responsible Entity Agfa Australia Pty Ltd
Reason/Issue Agfa Australia Pty Ltd has identified a potential safety issue related to the steel cable installed inside the 3.1 mobile column which supports the weight of the telescopic arm with the tube head.

The potential risk consists of an interference which occurs between the cable and a metal plate of the arm’s carriage due to the shape of the cable’s crimped terminals.

No incident related to the potential defect has been reported.
Recall Action Product Defect Correction
Recall Action Instructions Agfa Australia will inspect the cable of customer's device at the earliest opportunity and will replace it in case of any potential malfunction of the safety system.