Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00165-1
Product Name/Description AIRFLOW PROPHYLAXIS MASTER or AIRFLOW One

WATER Bottle
EG-121
Expiry date: 04/2026, 07/2026, 10/2026

NIGHT CLEANER Bottle
EG-120
Expiry date: 07/2026, 10/2026

PIEZON Bottle
EG-111
Expiry date: 04/2026, 07/2026

ARTG 221348
(EMS Oceania Pty Ltd- Dental abrasive air jet system)
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 27/03/2024
Responsible Entity EMS Oceania Pty Ltd
Reason/Issue EMS Oceania Pty Ltd has identified batches of bottle components of the AIRFLOW PROPHYLAXIS MASTER or AIRFLOW One Dental abrasive air jet systems that have not been manufactured to quality standards.

Root cause of the problem is found to be a weakness of the part above the ultrasound welding of the water bottle. This weakness is due to a higher concentration of black colorant in the bottle lid.

10 incidents have been reported as defective so far (0.092 %) within more than 1 year of use. No injuries were reported. Damage to the ceiling of the dental office were reported.
Recall Action Recall
Recall Action Instructions Cease use with immediate effect and return the bottles with the expiry date listed to the distributor.

This action has been finalised on 27/6/2024.