Type of Product |
Medical Device |
TGA Recall Reference |
RC-2024-RN-00144-1 |
Product Name/Description |
Anti-k. An in vitro diagnostic medical device (IVD)
Catalogue number: 007251
Expiry: 30/09/2024
Product UDI: 07611969001477
ARTG 217696 (Bio-Rad Laboratories Pty Ltd - Immunohaematology blood grouping antisera IVDs) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
22/02/2024 |
Responsible Entity |
|
Reason/Issue |
An anomaly has been identified that could pose a risk for patients. It has been detected that the anti-k may result in false positive reactions with k negative samples.
Cases reported thus far have been confirmed during our investigations. False positive reactions up to a “++” reaction may occur. |
Recall Action |
Product Defect Correction |
Recall Action Instructions |
Bio-Rad is requesting that customers affected by this notice take the following action in case you obtain reaction = “++” for anti-k well: - Use alternative method to confirm the result with other reagents available on the market, for example: o Tube testing with Seraclone Anti-k, Ref 808126, or o Other reagents intended to the identification of k (KEL2)
Note that Reactions = “+++” are true positives and indicate the presence of k (KEL2) antigen. |
Contact Information |
1800 224 354 - Bio-Rad Laboratories Technical Support |