Type of Product |
Medical Device |
TGA Recall Reference |
RC-2024-RN-00112-1 |
Product Name/Description |
Wrench Devices
Wrench, Cannulated Socket 10mm Part Number: 910-CW-10L UDI: 0084113214296
Wrench, Combo 10mm Part Number: 915-CW-10 UDI: 0084113214297
Wrench, Swivel 10mm Part Number: 915-FW-10 UDI: 0084113214300
Orthex Wrench 7mm Part Number: C915-CW-07 UDI: 0084113217782
ARTG 302175 (Orthopediatrics AUS Pty Ltd - Surgical procedure kit, orthopaedic, reusable) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
1/03/2024 |
Responsible Entity |
|
Reason/Issue |
OrthoPediatrics received nine complaints in relation to items without manufacturer logos imprinted. After investigation, OrthoPediatrics has identified these wrenches may corrode after multiple sterilisation cycles.
There is a remote possibility of a delay in surgery. Visual inspection of device would determine any corrosion present.
No adverse events have been reported to date. |
Recall Action |
Recall |
Recall Action Instructions |
The affected product must be returned to Orthopediatrics. Replacement product will be provided for device part numbers 910-CW-10L, 915-CW-10 and C915-CW-07. Part Number 915-FW-10 will no longer be available. Customers are to contact customer service to arrange return of the affected product. No additional corrective actions need to be taken for product already implanted. |
Contact Information |
jbatts@orthopediatrics.com - OrthoPediatrics |