Type of Product |
Medical Device |
TGA Recall Reference |
RC-2024-RN-00062-1 |
Product Name/Description |
Slishman Traction Splint (Gen 2)
Manufactured 2nd May 2023 Serial Numbers: G2-2023-0056 through to G2-2023-0075
ARTG 216497 (Tacmed Australia Pty Ltd - Splint, ) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
7/02/2024 |
Responsible Entity |
|
Reason/Issue |
After repeated use of the device the cord used to apply traction may be excessively worn down while resetting the device and subsequently break during application.
This problem could lead to the inability to apply traction via mechanical advantage and cause pain to the patient. |
Recall Action |
Recall |
Recall Action Instructions |
Customers are to immediately inspect and quarantine any affected stock.
Customers are to return affected stock to TacMed Australia, and replacement stock will be provided. |
Contact Information |
1300 862 633 - TacMed Australia |