Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00062-1
Product Name/Description Slishman Traction Splint (Gen 2)

Manufactured 2nd May 2023
Serial Numbers: G2-2023-0056 through to G2-2023-0075

ARTG 216497
(Tacmed Australia Pty Ltd - Splint, )
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/02/2024
Responsible Entity Tacmed Australia Pty Ltd
Reason/Issue After repeated use of the device the cord used to apply traction may be excessively worn down while resetting the device and subsequently break during application.

This problem could lead to the inability to apply traction via mechanical advantage and cause pain to the patient.
Recall Action Recall
Recall Action Instructions Customers are to immediately inspect and quarantine any affected stock.

Customers are to return affected stock to TacMed Australia, and replacement stock will be provided.