Type of Product |
Medical Device |
TGA Recall Reference |
RC-2023-RN-01100-1 |
Product Name/Description |
Stericlin Flat Roll 150mmx200m - 800304
Batch number: 3050168
Lot number: 4046-4145
ARTG 135955 (Med-Con Pty Ltd - Sterilization packaging, single use) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
21/12/2023 |
Responsible Entity |
|
Reason/Issue |
One of the two sealing side seams are missing over a length of about 25cm. All three grooves of the groove sealing are affected on the above-mentioned length.
The incompleteness of the seal is visually recognisable, and the seal seam is open or will open at a very slight stress level. The defective seal is repeating in the affected reels regularly roughly every meter within the roll.
The issue will cause the roll to not seal correctly and the product to not sterilize. Resulting in the failure in the sterilization process. Stericlin Flat roll 150mmx200m will not sterilize the instruments resulting in possible contamination.
This recall does not affect any other batches/lots/versions of #800304 – Stericlin Flat Roll 150mmx200m or any other Med-con products. This Batch 3050168 – Lot: 4046-4145 has been distributed to hospitals and medical distribution companies from January 2023. |
Recall Action |
Recall |
Recall Action Instructions |
Inspect stock immediately and quarantine affected stock with affected batch numbers on hand to prevent further use.
Return affected stock on hand to: Store 4/415 New Dookie Rd, Lemnos, VIC 3631
This action has been closed out on 21/3/2024. |
Contact Information |
03 5822 6600 - Med-Con |