Product Name/Description |
Sterile Water Syringe supplied with Atrium Express Dry Suction Dry Seal Chest Drain
Product Code: 4000-100N
Lot numbers: 466951, 467193, 467195, 469918, 469920, 471069, 472581, 474076, 474950, 475228, 475487, 477950, 485229, 485230, 487808, 489877, 490762, 492644, 496207, 496692, 497139, 498578.
ARTG 119726 (Getinge Australia Pty Ltd - Drainage system, pleural) |
Reason/Issue |
The Express chest drain was shipped with sterile water from Nurse Assist, which could be used to determine if an active pneumothorax (“air leak”) is present.
On November 8, 2023, Getinge received notice from Nurse Assist, LLC that its Sterile Water, USP, 30mL syringes were being recalled because they could not be verified to be sterile. Pre-packaged with every Express chest drain, the 30mL sterile water syringe is intended to fill the air leak monitor chamber for air leak monitor chamber for air leak detection during or after initial device set-up, if desired. Express Chest Drains labelling states that the fluid is sterile.
During a chest drain knock-over event (device not kept in upright position), the water in the air leak monitor chamber could migrate from the air leak monitor chamber to the drainage fluid collection chamber. The patient could potentially be exposed to an infectious pathogen; therefore, the potential risk to the patient is high. |