Reason/Issue |
Starting on 10 January 2024, the clock parameter used by the Model 2892 software application on the Model 3120 ZOOM programmer will display certain diagnostic dates incorrectly, with a year in the 1990s. The Model 2892 software application supports the remaining ALTRUA, INSIGNIA and family of pacemakers that were distributed between October 2001 and August 2018.
Affected diagnostic dates that will display incorrectly on or after 10 January 2024, include date of last battery test, last programmed date, last interrogation date, last reset counter/histograms, and the dates of follow-up tests such as intrinsic amplitude, pacing impedance, and pacing threshold tests.
The replacement indicator dates and date of events in the arrhythmia logbook are not affected and display correctly.
There is no impact to the performance of the pacemaker, actual longevity remaining of the pacemaker, nor battery gas gauge display. |