Type of Product Medical Device
TGA Recall Reference RC-2023-RN-01030-1
Product Name/Description Zoom Programming System

Model 2892 Software application on Model 3120 Zoom Programmer

ARTG 123599
(Boston Scientific Pty Ltd - ZOOM LATITUDE Programming System - Pacemaker programmer)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/12/2023
Responsible Entity Boston Scientific Pty Ltd
Reason/Issue Starting on 10 January 2024, the clock parameter used by the Model 2892 software application on the Model 3120 ZOOM programmer will display certain diagnostic dates incorrectly, with a year in the 1990s. The Model 2892 software application supports the remaining ALTRUA, INSIGNIA and family of pacemakers that were distributed between October 2001 and August 2018.

Affected diagnostic dates that will display incorrectly on or after 10 January 2024, include date of last battery test, last programmed date, last interrogation date, last reset counter/histograms, and the dates of follow-up tests such as intrinsic amplitude, pacing impedance, and pacing threshold tests.

The replacement indicator dates and date of events in the arrhythmia logbook are not affected and display correctly.

There is no impact to the performance of the pacemaker, actual longevity remaining of the pacemaker, nor battery gas gauge display.
Recall Action Product Defect Correction
Recall Action Instructions Customers are to follow the instructions in the customer letter (supplied to affected customers by Boston Scientific).

When using the Model 2892 software application on the Model 3120 ZOOM programmer, if a date includes a year in the 1990s
- Determine a precise date by interrogating an ALTRUA, INSIGNIA pacemaker with a Model 3300 LATITUDE programmer: or
- Determine an approximate date (within a two-day range) using the table in Appendix A.

Boston Scientific does not recommend prophylactic replacement of pacemakers due to incorrect dates.
Contact Information 02 8063 8340 - Boston Scientific