Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2023-RN-00948-1
Product Name/Description HeartStart Intrepid Monitor/Defibrillator

Product code: 867172

ARTG 281589
(Philips Electronics Australia Ltd - Physiologic-monitoring defibrillation system)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/11/2023
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue The current HeartStart Intrepid Service Manual does not detail IEC Class I electrical safety test verification methods. The lack of this testing does not introduce a failure, but it may prevent a failure from being detected. This could occur when an IEC Class I device is tested as a Class II device after a servicing event.

The issue was identified as the result of a service provider’s inquiry. There have been no reports of patient harm.
Recall Action Product Defect Correction
Recall Action Instructions Customers are advised that:
- The HeartStart Intrepid Monitor/Defibrillator may continue to be used if: the Class I device that has been opened and serviced is tested using the updated Service Manual which will be supplied to affected customers. A hard copy of the Service Manual Addendum will be supplied to customers along with the customer service letter. For customers using a CD version of the manual, they are to follow the hardcopy version provided.

Philips will contact customers to arrange complimentary retesting to Class I as necessary.
Contact Information 1800 251 400 - Philips Service Delivery Team