Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2022-RN-01316-1
Product Name/Description HeartWare (HVAD) System

Model number: 1104

Serial number: HW35315

ARTG 181875
(Medtronic Australasia Pty Ltd - Heartware Ventricular Assist System - Implantable ventricular circulatory assist system)
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 21/10/2022
Responsible Entity Medtronic Australasia Pty Ltd
Reason/Issue Medtronic is providing a letter as a follow-up to previous "Urgent Hazard Alert” RC-2020-RN-01496-1.

There are no new HVAD devices identified, Medtronic is sending this communication to all clinicians with patients currently on support.

Medtronic is providing additional recommendations for all patients on support (including subgroup 1 and 2).
Recall Action Hazard Alert
Recall Action Instructions Medtronic recommands the following as outlined in the customer letter (will be provided only to the impacted customers by Medtronic):

- Surgeons are advised to notify implanted customers to follow the additional Patient Management Recommendations provided in the customer letter.
- Reinforce the current Instructions for Use to patients and staff;

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.
Contact Information 0478 486 782 - Arthur - Medtronic