TGA Recall Reference RC-2020-RN-00033-1
Product Name/Description Green Rusch Lite Disposable Metal Laryngoscope Blade

Product Code: 004551004

Lot Number: 1812341

ARTG 286313
(Teleflex Medical Australia Pty Ltd - Laryngoscope Blade, Rigid, Single Use)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/01/2020
Responsible Entity Teleflex Medical Australia Pty Ltd
Reason/Issue Teleflex have received 11 customer complaints reporting that the light guide of the Rusch GreenLite Mac laryngoscope broke at the point of a welded joint, either prior to use during setup or during use inside the patient's mouth. The occurrence rate associated with this failure is approximately 0.0031%.

If this product failure occurs and is not identified prior to use, the consequences may include:

1) Delay in procedure while medical personnel retrieve the broken component and potentially locate an alternative laryngoscope to reintubate and proceed with tracheal intubation;
2) If the broken component cannot be located immediately medical personnel may require diagnostic imaging to assist in the retrieval process. Additional medical intervention (e.g., a bronchoscopy) and increased patient monitoring may also be required.
Recall Action Recall
Recall Action Instructions Customers are advised to immediately check inventory and quarantine affected stock.

Customers are advised to contact Teleflex Customer Services to organise return of affected goods. Teleflex will issue a credit note upon receipt of the returned product.

This action has been closed out on 04/01/2021