Print version of this report
TGA Action ID RC-2026-RN-00955-1
Type of Product Medical Device
Product Name/Description NeuroWorks/SleepWorks 10.0.0 Software (INTL) used with SD PLUS FLEXI amplifier

NeuroWorks / SleepWorks Software
Part Number: 041906

SD PLUS FLEXI Amplifier
Part Number: PK1288 (USB), PK/EXP1290 (NET)
ARTG Number 197550
Unique Device Identifier (UDI) 00382830061759
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Natus has become aware of complaints related to display differences when monitoring live recordings. A complaint was received where the live EEG shown on a remote monitoring station differed from the live EEG on the acquisition computer and from the same study was subsequently reviewed. This affects only NeuroWorks / SleepWorks 10.0.0 used with the SD PLUS FLEXI amplifier.

The investigation has determined that the problem is associated with remote monitoring of studies recorded with NeuroWorks / SleepWorks 10.0.0 and with SD PLUS FLEXI amplifier. The waveforms displayed on the remote monitoring station have incorrect amplitudes.

Due to this discrepancy, users may receive a misleading representation of the patient's EEG during remote monitoring. The defect only occurs when remote monitoring using a monitoring or review station.
Instructions for Customers/Users Natus will provide a software update to correct the problem.

In the interim, customers may continue to use their device, however they must immediately cease the use of remote monitoring when using NeuroWorks / SleepWorks 10.0.0 software with the SD PLUS FLEXI amplifier. They can continue recording and reviewing with NeuroWorks / SleepWorks 10.0.0 software with the SD PLUS FLEXI amplifier.
Action Commencement Date 19/11/2026
Responsible Entity Natus Neuro Australia Pty Ltd
Contact Information 1800 290 772 or FCA@Natus.com - Natus Neuro Australia