TGA Action ID RC-2026-RN-00949-1
Type of Product Medical Device
Product Name/Description Instrument Trays

TS Universal Baseplate Finishing Tray, Triathlon
Catalogue Number: 6543-6-506
Lot Numbers: 19114556, 200009859

TS Femoral Prep Tray, Triathlon
Catalogue Number: 6543-6-507
Lot Numbers: 200009873, 20447278

TS Misc. Instruments Tray, Triathlon
Catalogue Number: 6543-6-519
Lot Numbers: 19114564, 200009912

Triathlon Hinge Femoral Prep Tray
Catalogue Number: 6543-6-514
Lot Numbers: 19114388, 70945430, 70945448
ARTG Number 285535
Unique Device Identifier (UDI) 07613327548082, 07613327557138, 07613327555127, 07613327562859
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Stryker has identified that the reprocessing instructions provided in LSTPI-B specify a sterilisation dry time of 30 minutes. In fact, these trays require a sterilisation dry time of 40 minutes when sterilised in blue wrap. As a result, when blue wrap is used, the 30 minute dry time may lead to a small amount of residual moisture remaining after sterilisation.

There is a potential pathway for contamination of the tray if moisture is present. In the case that the set is recognised as containing moisture prior to or during surgery, this would result in a surgical delay. It may be as short as 10-15 minutes if another sterile set is available. If no backup set is available and the set is identified as unsterile, this could result in case cancellation.
Instructions for Customers/Users Customers are to:
1. Review and follow updated instructions provided in LSTPI-B Cleaning, Sterilization, Inspection, and Maintenance of Reusable Medical Devices Reference Guide.
Action Commencement Date 19/11/2026
Responsible Entity Stryker Australia Pty Ltd