| TGA Action ID |
RC-2026-RN-00949-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Instrument Trays
TS Universal Baseplate Finishing Tray, Triathlon Catalogue Number: 6543-6-506 Lot Numbers: 19114556, 200009859
TS Femoral Prep Tray, Triathlon Catalogue Number: 6543-6-507 Lot Numbers: 200009873, 20447278
TS Misc. Instruments Tray, Triathlon Catalogue Number: 6543-6-519 Lot Numbers: 19114564, 200009912
Triathlon Hinge Femoral Prep Tray Catalogue Number: 6543-6-514 Lot Numbers: 19114388, 70945430, 70945448 |
| ARTG Number |
285535 |
| Unique Device Identifier (UDI) |
07613327548082, 07613327557138, 07613327555127, 07613327562859 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Stryker has identified that the reprocessing instructions provided in LSTPI-B specify a sterilisation dry time of 30 minutes. In fact, these trays require a sterilisation dry time of 40 minutes when sterilised in blue wrap. As a result, when blue wrap is used, the 30 minute dry time may lead to a small amount of residual moisture remaining after sterilisation.
There is a potential pathway for contamination of the tray if moisture is present. In the case that the set is recognised as containing moisture prior to or during surgery, this would result in a surgical delay. It may be as short as 10-15 minutes if another sterile set is available. If no backup set is available and the set is identified as unsterile, this could result in case cancellation. |
| Instructions for Customers/Users |
Customers are to: 1. Review and follow updated instructions provided in LSTPI-B Cleaning, Sterilization, Inspection, and Maintenance of Reusable Medical Devices Reference Guide. |