| TGA Action ID |
RC-2026-RN-00929-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Rivacor 5 and Rivacor 7 ICDs and CRT-Ds
Multiple models and serial numbers
ARTG's: 396015, 396011, 396217, 396264, 396251, 326286 (cancelled) |
| ARTG Number |
396012, 396013, 396045, 396233, 396258 |
| Unique Device Identifier (UDI) |
Attached |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class I |
| Reason for the Action |
BIOTRONIK has become aware of an increased likelihood of premature battery depletion due to lithium plating in a subset of Rivacor 5 and Rivacor 7 implantable cardioverter defibrillators (ICDs) and cardiac resynchronisation therapy defibrillators (CRT-Ds). Lithium plating is a rare phenomenon that may cause a battery drain at a higher rate than normal.
The current action is an extension of the advisory population previously notified in March 2021 (RC-2021-RN-00667-1).
These devices were distributed from September 2018 to March 2022. Please note that NO other BIOTRONIK ICDs or CRT-Ds are affected by this problem.
To date, no serious injuries or deaths have been reported in association with the affected Rivacor subset. Most reported cases have resulted in an earlier-than-expected need for device replacement due to reduced battery longevity.
The current cumulative observed rate of confirmed premature battery depletion due to lithium plating in the affected Acticor/Rivacor device subset is 0.68%. The earliest observed onset for devices experiencing this problem is 27 months. |
| Instructions for Customers/Users |
Customers should: - Note that NO affected devices remain available for implantation.
- Follow-up: Continue with the standard patient follow-up schedule. During in-office or Home Monitoring follow-ups, verify the status of the device and battery. Please inform your BIOTRONIK representative if you observe any unusual device behaviour.
- Home Monitoring: Use Home Monitoring when possible. It provides timely ERI notifications and has reduced the probability of loss of shock therapy by 57% and loss of pacing therapy by 82% in confirmed cases. BIOTRONIK will provide CardioMessenger devices free of charge to monitor implants affected by this urgent Product Alert.
- Unexpected ERI notification: For a device subject to this urgent product alert, a timely replacement should be considered based on the patient's underlying condition: a. For patients who are not pacemaker dependent, or patients with a primary prevention ICD, device replacement within one week after ERI notification is recommended. b. For pacemaker-dependent patients, device replacement is recommended as soon as possible after ERI notification.
- BIOTRONIK does not recommend prophylactic replacement. |
| Action Commencement Date |
19/11/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 9497 3700 - BIOTRONIK Customer Care |