TGA Action ID RC-2026-RN-00929-1
Type of Product Medical Device
Product Name/Description Rivacor 5 and Rivacor 7 ICDs and CRT-Ds

Multiple models and serial numbers

ARTG's: 396015, 396011, 396217, 396264, 396251, 326286 (cancelled)
ARTG Number 396012, 396013, 396045, 396233, 396258
Unique Device Identifier (UDI) Attached
Action Type Product Alert
Action Level Hospital
Hazard Classification Class I
Reason for the Action BIOTRONIK has become aware of an increased likelihood of premature battery depletion due to lithium plating in a subset of Rivacor 5 and Rivacor 7 implantable cardioverter defibrillators (ICDs) and cardiac resynchronisation therapy defibrillators (CRT-Ds). Lithium plating is a rare phenomenon that may cause a battery drain at a higher rate than normal.

The current action is an extension of the advisory population previously notified in March 2021 (RC-2021-RN-00667-1).

These devices were distributed from September 2018 to March 2022. Please note that NO other BIOTRONIK ICDs or CRT-Ds are affected by this problem.

To date, no serious injuries or deaths have been reported in association with the affected Rivacor subset. Most reported cases have resulted in an earlier-than-expected need for device replacement due to reduced battery longevity.

The current cumulative observed rate of confirmed premature battery depletion due to lithium plating in the affected Acticor/Rivacor device subset is 0.68%. The earliest observed onset for devices experiencing this problem is 27 months.
Instructions for Customers/Users Customers should:
- Note that NO affected devices remain available for implantation.

- Follow-up: Continue with the standard patient follow-up schedule. During in-office or Home Monitoring follow-ups, verify the status of the device and battery. Please inform your BIOTRONIK representative if you observe any unusual device behaviour.

- Home Monitoring: Use Home Monitoring when possible. It provides timely ERI notifications and has reduced the probability of loss of shock therapy by 57% and loss of pacing therapy by 82% in confirmed cases. BIOTRONIK will provide CardioMessenger devices free of charge to monitor implants affected by this urgent Product Alert.

- Unexpected ERI notification: For a device subject to this urgent product alert, a timely replacement should be considered based on the patient's underlying condition:
a. For patients who are not pacemaker dependent, or patients with a primary prevention ICD, device replacement within one
week after ERI notification is recommended.
b. For pacemaker-dependent patients, device replacement is recommended as soon as possible after ERI notification.

- BIOTRONIK does not recommend prophylactic replacement.
Action Commencement Date 19/11/2026
Responsible Entity Biotronik Australia Pty Ltd
Contact Information (02) 9497 3700 - BIOTRONIK Customer Care