Print version of this report
TGA Action ID RC-2026-RN-00928-1
Type of Product Medical Device
Product Name/Description Cardiosave Intra-Aortic Balloon Pumps (IABP) Power Management Board and Solenoid Control Boards

Cardiosave Hybrid IABP
Product Codes: D0998-00-0800-31, D0998-00-0800-32, D0998-00-0800-35, D0998-00-0800-52, D0998-00-0800-55, D0998-UC-0800-55

Cardiosave Rescue IABP
Product Codes: D0998-00-0800-83

Manufacturing Dates: May 1, 2011 to July 31, 2026
ARTG Number 118266
Action Type Product Correction
Action Level Hospital
Hazard Classification Class I
Reason for the Action Getinge has identified that some Cardiosave devices may contain a Power Management Board (PMB) and/or Solenoid Control Board (SCB) with one or more capacitors installed with reversed polarity.

These capacitors may degrade over time, potentially causing Cardiosave device battery charging failures, an inability to power on, interruption of therapy, or unexpected shutdown.

Users cannot identify the affected capacitors during normal operation.

As of October 30, 2026 Getinge has received two (2) complaints related to this problem. No associated deaths or serious injuries have been reported.
Instructions for Customers/Users The Cardiosave device may continue to be used to provide therapy within the hospital environment. However,
· Until the affected Cardiosave device inspection and/or correction is completed, make sure an alternate IABP is available, or there are alternate means of supporting the patient.
· It is recommended that Cardiosave not be used for interhospital transport, particularly for high acuity patients, until the affected IABP is inspected and any necessary work has been completed.
· Do not attempt to inspect or replace the PMB or SCB. A Getinge Service Representative will inspect the installed board serial numbers and replace any affected board.

If an unexpected shutdown occurs:
· Follow the troubleshooting steps and interventions as directed in the Cardiosave device Instructions for Use (IFU) and help screen.
· If you cannot resume therapy, identify an alternate IABP console to resume therapy or select other therapies as the patient’s condition warrants.
· If you cannot resume therapy with the Cardiosave IABP, remove the pump from service and contact your Getinge Service Representative for additional assistance.
Action Commencement Date 19/11/2026
Responsible Entity Getinge Australia Pty Ltd
Contact Information 1800 438 464 - Getinge Customer Service