TGA Action ID RC-2026-RN-00816-1
Type of Product Medical Device
Product Name/Description Abre Venous Self-Expanding Stent System

All Model Numbers

All Lot Numbers
ARTG Number 448908
Unique Device Identifier (UDI) See appendix A of the letter
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Medtronic is updating the Instructions for Use (IFU) for the Abre Venous Self-Expanding Stent System. This update provides a validated manual deployment work around method to help mitigate potential harms in the rare event they experience a partial stent deployment.

As of 14 September 2026, Medtronic reported 31 partial stent deployments (0.015%): 27 caused no or negligible harm, while four required further intervention or treatment.

There are no product retrievals or disposals requested by Medtronic as part of this IFU update.
Instructions for Customers/Users Customers are asked to review the upcoming updates to the IFU in the customer letter and share this notice with all those who need to be aware.

A link to a video demonstration of the Deployment Workaround Method can befound in the customer letter.
Action Commencement Date 13/10/2026
Responsible Entity Medtronic Australasia Pty Ltd