TGA Action ID RC-2026-RN-00501-1
Type of Product Medical Device
Product Name/Description Hotline Fluid Warming Set

Affected List Number: L-70

Affected Lot Numbers: 4293569, 4326087 & 4337725
ARTG Number 398170
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action ICU Medical received seven complaints involving an increased likelihood of leakage in three lots of L-70 HOTLINE Fluid Warming Sets. This problem is due to potentially inadequate solvent application at the tubing to return connector bond, with leakage occurring from an inadequate bond at this connection.

Delay of therapy may occur if a L-70 set is deemed unusable during pre-use checks, and a replacement set is not immediately available. Interruption of therapy may occur if leakage during use requires replacement of the set. Under-delivery of medication, exposure to infectious or toxic agents, and air in line, are also possible due to leakage in sets.
Instructions for Customers/Users Customers are to
- Inspect inventory and quarantine affected lots of HOTLINE Fluid Warming Sets
- Complete the Customer Response Form as instructed in the customer letter, and destroy the product in accordance with institutional procedures
Action Commencement Date 1/07/2026
Responsible Entity Smiths Medical Australasia Pty Ltd
Contact Information 1800 654 949 - Customer Service