Print version of this report
TGA Action ID RC-2025-RN-00014-1
Type of Product Medical Device
Product Name/Description MCL31 DERMABLATE & JULIET HANDPIECE WITH STERISPOT COVERS

MCL31 DERMABLAT
Serial No: 182000729, 182000745, 182000747, 182000786, 182000811, 182000865, 182000867, 182000873, 182000895, 182000954, 182001005, 182001135 182001595, 182001628

JULIET HANDPIECE
Serial No: 390100120, 390100163, 390100187, 390100203, 390100205, 390100228, 390100247, 390100304, 390100306, 390100309, 390100347, 390100477, 390100528, 390100653, 390100786, 390100790

ARTG 234391
ARTG Number 234391
Action Type Recall
Action Level Retail
Hazard Classification Class II
Reason for the Action The TGA has completed a post-market review of all energy-based devices used for vaginal rejuvenation included in the Australian Register of Therapeutic Goods (ARTG). The review found there was insufficient clinical evidence to support the therapeutic use and long-term safety of these devices.

Claims related to genitourinary indications and the use of associated handpieces have been removed from the Instructions for Use (IFU) for the MCL31 Dermablate device.

The Juliet handpiece with SteriSpot covers, which was specifically for genitourinary indications has been withdrawn from the market.

The MCL31 Dermablate device continues to be included on the ARTG and is not affected by the recall except for the amendment to the IFU. MCL31 Dermablate can continue to be used for all non-genitourinary applications as per the IFU and in accordance with the indications included on the ARTG.
Instructions for Customers/Users Recall:
Device Consulting will contact customers to arrange for return of the Juliet handpieces

Product Defect Correction:
Device Consulting will contact customers to replace the IFU customers currently have for the MCL31 Dermablate.

This action has been finalised on 21/7/2025.
Action Commencement Date 10/01/2025
Responsible Entity Device Consulting Pty Ltd
Contact Information 03 9998 2020 - Device Consulting