Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00238-1
Product Name/Description Pentax ED-3490TK Video Duodenoscope

ARTG Number: 194533
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/04/2016
Responsible Entity Hoya Lens Australia Pty Ltd
Reason/Issue Pentax is issuing updated, validated manual reprocessing instructions for the ED-3490TK Video Duodenoscope to replace those provided in the original device labelling. The updated reprocessing instructions for the ED-3490TK Video Duodenoscope include a more rigorous protocol for pre-cleaning, manual cleaning, high-level disinfection and liquid chemical sterilisation procedures. In addition, updated instructions include additional text, figures, cautions and warnings intended to clarify the validated reprocessing procedure.

If not properly reprocessed, residual body fluids and organic debris may remain in microscopic crevices of the device following an attempted cleaning and high level disinfection.
Recall Action Recall for Product Correction
Recall Action Instructions Pentax is providing affected customers with updated Instructions for Use intended to clarify the validated reprocessing procedure. This action has been closed-out on 27/01/2017.
Contact Information 03 9390 6838 - Pentax Medical